Alerta De Seguridad para Commercial Name: Three-Dimensional Planning System for Eclipse Radiotherapy Technical Name: ANVISA Registration Number: 10405410007 Hazard Class: III Affected Model: Eclipse Serial Numbers Affected: H48J544

Según Agência Nacional de Vigilância Sanitária (ANVISA), este evento ( alerta de seguridad ) involucró a un dispositivo médico en Brazil que fue producido por Varian Medical Systems Brasil Ltda; Varian Medical Systems.

¿Qué es esto?

Las alertas proporcionan información importante y recomendaciones sobre los productos. Aunque se haya emitido una alerta, esto no significa necesariamente que el producto se considera peligroso. Las alertas de seguridad, dirigidas a trabajadores de la salud y a usuarios, pueden incluir retiro de equipos. Pueden ser escritas por los fabricantes, pero también por funcionarios del área de salud.

Más información acerca de la data acá
  • Tipo de evento
    Safety alert
  • ID del evento
    2165
  • Fecha
    2017-02-13
  • País del evento
  • Fuente del evento
    ANVISA
  • URL de la fuente del evento
  • Notas / Alertas
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notas adicionales en la data
    The problem does not occur when additional changes are made to outline structures. The rescue process is performed correctly after performing the additional steps below. 1) Enable / disable the visibility of the structure; 2) Change the color of the structure; 3) Approve structure; 4) Undo and redo the last operation; or 5) Modify the structure with a tool other than Draw Planar Contour, Brush, Draw Geometrical Shape, or Eraser After you edit a structure, perform any of the five steps above and save. Check and approve structures after completing targeting to avoid using incomplete structures in treatment planning. Ensure that a qualified physician verifies the accuracy and placement of all patient structures (target structure and critical structures) used for treatment planning and to evaluate plans prior to patient treatment. #### UPDATED ON 08/24/2017, the company submitted completion report of the field action proving the accomplishment of correction as planned.
  • Causa
    Contours are not consistently saved after editing this occurs on systems that have been upgraded to eclipse version 13.6 mr2 in just a few specific changes. modifications made in version 13.6mr2 for the contouring, smartadapt ™, and smartsegmentation ™ workspaces result in inconsistently saved contours in eclipse. the anomaly is displayed when all of the following conditions occur: 1. the contours of the structure are modified using the draw planar contour, brush, draw geometrical shape, or eraser tools, ; e 2. no additional tools are applied to the structure; e 3. structural visibility, approval status, and structure properties are not changed. when all these conditions occur, the structure modification is not recognized or saved by the software version 13.6mr2.
  • Acción
    Field Action Code CP-28454 triggered under the responsibility of the company Varian Medical Systems Brazil Ltda. Company will make correction in the field.