Alerta De Seguridad para COMMERCIAL NAME: VASCULAR STARCLOSE CLOSURE SYSTEM, REGISTER NUMBER: 80146501353, CLASS: IV - MAXIMUM RISK, LOTS: 43034-6H at 47139-6H

Según Agência Nacional de Vigilância Sanitária (ANVISA), este evento ( alerta de seguridad ) involucró a un dispositivo médico en Brazil que fue producido por PERCLOSE INC. (ABBOTT VASCULAR DEVICES).

¿Qué es esto?

Las alertas proporcionan información importante y recomendaciones sobre los productos. Aunque se haya emitido una alerta, esto no significa necesariamente que el producto se considera peligroso. Las alertas de seguridad, dirigidas a trabajadores de la salud y a usuarios, pueden incluir retiro de equipos. Pueden ser escritas por los fabricantes, pero también por funcionarios del área de salud.

Más información acerca de la data acá
  • Tipo de evento
    Safety alert
  • ID del evento
    856
  • Fecha
    2007-02-12
  • País del evento
  • Fuente del evento
    ANVISA
  • URL de la fuente del evento
  • Notas / Alertas
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notas adicionales en la data
    The implication of the quality deviation, that is, the release of the clip that does not occur in the correct place, is noticeable at the time of use of the product. The remedial action, if the problem is verified, is the use of manual compression. There are no long-term consequences on the use of the product. We imported 2 lots, total of 260 units, that fit within this range. The Company explains that there is no record of any contact at Abbott Laboratories Customer Service Center regarding the product. Of these 260 units, 05 were effectively used in Brazil. These units did not present a quality deviation and the patients follow well without any postoperative complications. Of the 260 units imported, 200 units of lot 45052 were not even distributed in the country and are already segregated for return to the manufacturer. Of the 60 units of lot 45053, 48 units are segregated to be returned to the manufacturer, the remaining 12 units were distributed as follows: - 07 units were used in medical demonstration (not used in patients); - 05 units were marketed. The UTVIG - Technovigilance Unit will be accompanying the entire return process.
  • Causa
    Growth in premature release rate of vessel localizer wings to somewhere around 0.4% (1 case in 250 units). this deviation was specifically verified in the batches range from 43034-6h to 47139-6h.
  • Acción
    Abbott made the voluntary decision to collect this batches range from 43034-6H to 47139-6H of the product for return to the manufacturer. Note: To access the attachments: http://www.anvisa.gov.br/tecnovigilancia/alertas/anexos_2007/856.pdf