Alerta De Seguridad para D-Fend / D-Fend + Moisture Condensers Used with E / MC (Ai) O (V) Compact Respiratory Gas Modules (CAR): CARDIOCAP 5, Hazard Class III, Record: 80071260203. X) and Cardiocap / 5 monitors. /////////. . Affected lots of moisture condensers :. D-Fend (Part Numbers): 876446-HEL: J9964985, J10201698, J10258248, J10284871, J10284872, J10344108. D-Fend + (Part Numbers): 881319-HEL: J9906966, J9965152, J10258251, J10300071 ////////////. . Products that can use E / MC (Ai) O (V) (X) Compact Respiratory Gas Modules and Cardiocap / 5 monitors are registered with ANVISA as: MONITOR OF VITAL SIGNS ::::: Product: Carescape Monitor, Model: B850 - Registro ANVISA: 80071260167 ////. Product: Patient Monitor, Model: Carescape B650- Register ANVISA: 80071260166 //// Product: Carescape Modular Monitor Model: B450 - ANVISA Register: 80071260333 //// Product: Cardiocap / 5 - ANVISA Register: 80071260203 //// Carescape Ventilator, Model: Carescape R860 - Registro ANVISA: 80071260334 //// Product: Ventilator for Intensive Care Engstrom Carestation - Registro ANVISA: 80071260230 //// ANESTHESIA APPARATUS ::::: Product : Anesthesia System Avance Model: Avance CS2 and Avance Pro - ANVISA Registration: 80071260306 //// Product: Anesthesia Device, Model: Aisys - Registro ANVISA: 80071260272

Según Agência Nacional de Vigilância Sanitária (ANVISA), este evento ( alerta de seguridad ) involucró a un dispositivo médico en Brazil que fue producido por GE Healthcare do Brasil; GE HEALTHCARE FINLAND OY.

¿Qué es esto?

Las alertas proporcionan información importante y recomendaciones sobre los productos. Aunque se haya emitido una alerta, esto no significa necesariamente que el producto se considera peligroso. Las alertas de seguridad, dirigidas a trabajadores de la salud y a usuarios, pueden incluir retiro de equipos. Pueden ser escritas por los fabricantes, pero también por funcionarios del área de salud.

Más información acerca de la data acá
  • Tipo de evento
    Safety alert
  • ID del evento
    1564
  • Fecha
    2015-04-20
  • País del evento
  • Fuente del evento
    ANVISA
  • URL de la fuente del evento
  • Notas / Alertas
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notas adicionales en la data
    According to the company, there is a risk that D-fend condenser units in the field are defective. These defective units can lead to delays in the measurement of CO2, O2 and anesthetic agents. The measurement of CO2 is important both in the Operating Rooms and Intensive Care Units (ICUs) so that the measurement of patient ventilation is possible. When this problem is detected by the equipment, the connected monitoring device will issue an audible alarm and will display the following warning messages: "low gas sample flow", "replace water trap" condenser) or "sampling line blocked". Ignored or prolonged alarms can lead to an adverse outcome if the CO2 value changes rapidly in certain clinical conditions. ### Update of the field action: UPDATED ON 10/23/2017, the company presented the report of completion of the field action proving the sending of the safety notice to the client with evidence of science and all actions completed.
  • Causa
    Occlusion, of specific batches, of the condensers of humidity d-fend / d-fend +. this failure may occur immediately after a replacement. occlusion of the moisture condenser can cause a loss of airway respiratory gas monitoring.
  • Acción
    The record holder recommends identifying the affected and unaffected products. The user must check the moisture capacitors through the batch number to identify the affected units. The batch number can be found in the cartons (white label in the box) and the individual moisture capacitors (white label on the front). Affected lots: D-Fend (Part Numbers): 876446-HEL: J9964985, J10201698, J10258248, J10284871, J10284872, J10344108 D-Fend + (Part Numbers): 881319-HEL: J9906966, J9965152, J10258251, J10300071 /// - Moisture condenser boxes with the batch numbers specified above without the Green Point Adhesive are affected. - Moisture condenser boxes with the batch numbers specified above with the Green Point Adhesive next to the batch number are not affected. If the lot is not affected, these products may remain in use / stock. / / In case the lot is affected, the user must: 1) Collect and discard all affected moisture capacitors. 2) Complete and return the Medical Device Correction Confirmation form attached to the Urgent Security Notice Letter. Detailed instructions SEE WARNING LETTER.

Manufacturer