Alerta De Seguridad para Equipment for angiography Artis, models Artis zee ceiling; Artis zee floor; Artis zeego, registration 10234230190, hazard class III, lots / series: 135297; 135329; 135785; 135816; 135825; 135835; 135848; 135869; 135941; 135969; 135980; 135984; 135991; 136056; 136060; 136068; 136115; 136119; 136136; 136144; 136199; 136213; 136214; 136276; 136282; 136283; 136284; 136285; 136288; 136301; 136320; 136355; 136363; 136369; 136373; 136419; 136421; 136424; 136425; 136440; 136443; 136457; 136492; 136580; 146646; 160143; 160343; 160360; 160467.

Según Agência Nacional de Vigilância Sanitária (ANVISA), este evento ( alerta de seguridad ) involucró a un dispositivo médico en Brazil que fue producido por Siemens Ltda.; SIEMENS AG..

¿Qué es esto?

Las alertas proporcionan información importante y recomendaciones sobre los productos. Aunque se haya emitido una alerta, esto no significa necesariamente que el producto se considera peligroso. Las alertas de seguridad, dirigidas a trabajadores de la salud y a usuarios, pueden incluir retiro de equipos. Pueden ser escritas por los fabricantes, pero también por funcionarios del área de salud.

Más información acerca de la data acá
  • Tipo de evento
    Safety alert
  • ID del evento
    1652
  • Fecha
    2015-07-24
  • País del evento
  • Fuente del evento
    ANVISA
  • URL de la fuente del evento
  • Notas / Alertas
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notas adicionales en la data
    Considering the problem related to the locking of the system in the "Fluoro Bypass" mode, it is possible that there is no radiation release and no fluoroscopic images are acquired. The risk would be more related to the delay due to the system restart. In the second problem presented the risk are related to the possible unexpected movement of the equipment resulting in collision.
  • Causa
    Two problems with the system have been described by the company that holds the registry: the first would be in case of an error in the equipment and it is in the "bypass fluoro" mode, while the x-ray blocking function is active, it will not be the only way out of this function is to solve the root cause of the equipment being in the "bypass fluoro" mode or to restart the system. the other problem is related to the possible failure of the position sensor on the articulated base axis that is not detected by the system software. following activation of operator movement, the c-arm system may move unexpectedly faster than normal.
  • Acción
    The company intends to continue to use the equipment until the company engineer's visit is made for field correction of the equipment. The field correction corresponds to the software update by installing two packages (11 and 12).

Manufacturer

  • Empresa matriz del fabricante (2017)
  • Source
    ANVSANVISA