Alerta De Seguridad para Ingenuity CT / Brilliance CT / Ingenuity Flex / MX8000 Dual v. EXP. Model: Brilliance CT (6, 10, 16, 16 Power, 40, 64, Big Bore, iCT, iCT SP - Registro Anvisa nº 10216710191, 10216710209 e 10216710142

Según Agência Nacional de Vigilância Sanitária (ANVISA), este evento ( alerta de seguridad ) involucró a un dispositivo médico en Brazil que fue producido por Philips Medical Systems.

¿Qué es esto?

Las alertas proporcionan información importante y recomendaciones sobre los productos. Aunque se haya emitido una alerta, esto no significa necesariamente que el producto se considera peligroso. Las alertas de seguridad, dirigidas a trabajadores de la salud y a usuarios, pueden incluir retiro de equipos. Pueden ser escritas por los fabricantes, pero también por funcionarios del área de salud.

Más información acerca de la data acá
  • Tipo de evento
    Safety alert
  • ID del evento
    1406
  • Fecha
    2014-04-14
  • País del evento
  • Fuente del evento
    ANVISA
  • URL de la fuente del evento
  • Notas / Alertas
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notas adicionales en la data
    There is a risk of serious injury to the patient and Technician / Operator as a result of table movement. For the patient: • There is a possibility of withdrawal or movement of invasive medical devices, such as IV, tracheostomy and tracheostomy drapes. To the Technician / Operator: • There is a possibility of crushing point in the subframe; and • A possibility of locking between the table and the Gantry if the Technician / Operator remains between them to move the patient ahead
  • Causa
    If the coupling is not properly locked during operation of the device, the table frame moves freely, causing accidental horizontal displacement. it results in serious damage with probability of occurrence not expected. risk analysis: acceptable.
  • Acción
    Sending User Safety Warning

Manufacturer

  • Empresa matriz del fabricante (2017)
  • Source
    ANVSANVISA