Alerta De Seguridad para Integris Monitor Arms

Según Agência Nacional de Vigilância Sanitária (ANVISA), este evento ( alerta de seguridad ) involucró a un dispositivo médico en Brazil que fue producido por Hill-Rom Co Inc A Hillenbrand Industry.

¿Qué es esto?

Las alertas proporcionan información importante y recomendaciones sobre los productos. Aunque se haya emitido una alerta, esto no significa necesariamente que el producto se considera peligroso. Las alertas de seguridad, dirigidas a trabajadores de la salud y a usuarios, pueden incluir retiro de equipos. Pueden ser escritas por los fabricantes, pero también por funcionarios del área de salud.

Más información acerca de la data acá
  • Tipo de evento
    Safety alert
  • ID del evento
    365
  • Fecha
    2002-01-16
  • País del evento
  • Fuente del evento
    ANVISA
  • URL de la fuente del evento
  • Notas / Alertas
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notas adicionales en la data
    UNTIL THIS DATE DOES NOT CONSERT PRODUCT REGISTRATION IN BRAZIL
  • Causa
    The brackets of this monitor are associated with 03 potential risks: (1) when the arm mounting pins loosen, the monitor may fall off the arm. this risk can occur when the professional tries to move the monitor horizontally or vertically. (2) mechanical failure of the arm mounting inner knee may cause the monitor to fall when it comes from the lower position to the rest position. this risk can occur when the professional tries to turn the arm vertically. (3) if the mounting was improperly installed the monitor may fall from the stand. this risk can occur when the professional tries to move the monitor horizontally or vertically. any of these problems can cause injury to the patient or operator or damage to the monitor and the arm of the patient. the manufacturer initiated a correction by an urgent health product correction letter dated december 4, 2001.
  • Acción
    Make sure you received the Health Product Urgent Correction letter dated December 4, 2001 and the 0.20-inch-inch pin-gauge pin from Hill-Rom. The manufacturer requires you to discontinue use of any monitor arm in your inventory until Hill-Rom's employee / inspector staff inspects and verifies that the unit is properly mounted. If it is essential to continue using the product Hill-Rom recommends that you notify the practitioner of the potential hazards described above, position the monitor as far as possible from the area where the patient will be and perform the inspection of the monitor arm. To inspect the monitor arm use the pin to check for clearance between the bottom of the base and the rotating adapter. If the pin comes fully into either side of the swivel adapter, remove the monitor arm from its service immediately. The manufacturer requests that you send the copy of the Urgent Correction of Health Products letter to all relevant personnel. In theory, you should have already been contacted by Hill-Rom regarding this correction. If this did not happen or you need more information, contact your local representative.