Alerta De Seguridad para On Board Imager / VARIAN RADIOTHERAPY SYSTEM with CBCT, Versions 1.4, 1.5 and 1.6 with CBCT 2.1.05 SP1-SP6, registration 10405410011, risk class III, lots / series H074200; H073676; H074373; H074390; H070897; H071472; H071657; H070418; H074494; H073721; H073249; H072087; H071455; H074151; H074298.

Según Agência Nacional de Vigilância Sanitária (ANVISA), este evento ( alerta de seguridad ) involucró a un dispositivo médico en Brazil que fue producido por Varian Medical Systems Brasil Ltda.; Varian Medical Systems INC.

¿Qué es esto?

Las alertas proporcionan información importante y recomendaciones sobre los productos. Aunque se haya emitido una alerta, esto no significa necesariamente que el producto se considera peligroso. Las alertas de seguridad, dirigidas a trabajadores de la salud y a usuarios, pueden incluir retiro de equipos. Pueden ser escritas por los fabricantes, pero también por funcionarios del área de salud.

Más información acerca de la data acá
  • Tipo de evento
    Safety alert
  • ID del evento
    1830
  • Fecha
    2016-02-29
  • País del evento
  • Fuente del evento
    ANVISA
  • URL de la fuente del evento
  • Notas / Alertas
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notas adicionales en la data
    Reinforced the need to adopt the following instructions: OBI version 1.5 and OBI 1.6: Navigate to the OBI application and select the "Cancel" button. The cancellation process can take anywhere from 5 to 30 seconds to complete. Restart the CBCT * acquisition process (See document CP-22609 in its entirety). OBI Version 1.4: Disable the CBCT field on the 4DITC workstation. Restart the OBI computer (s) and Reconstructor. Restart the OBI application. Enable the CBCT field on the 4DITC workstation. All versions of OBI: Contact your local Varian Medical Systems service department to report an incident of CBCT acquisition failure. #### UPDATED ON 08/24/2017, the company sent a completed action report of the field proving that a correction has been made as expected.
  • Causa
    Varian has received numerous reports on the locking of the acquisition of the on-board imager® cone beam ct [cbct] that prevents the reconstruction of the image. a repeat of the scan is required. there are two possible symptoms: 1) the cbct application stops updating the display during acquisition. 2) an exception occurs during acquisition. although the cbct projections have been acquired, it will not be possible to generate the reconstructed cbct and the scan must be repeated. varian company describes that no administration failures or injuries were reported due to this problem.
  • Acción
    Action code CP-22609. Send security warning to all affected clients.