Alerta De Seguridad para OPTIUM XCEED DIABETES MONITORING SYSTEM - OPTIUM XCEED DIABETES CONTROL SYSTEM - Record no. 80146501194 - ListNo .: 70099-01, 70098-01, 70098-01. Lots: 01K180D, 01K124D, 02K102D, 02K107D, 02K127D, 03K123D, 07K199D, 05K204D, 05K204D, 06K179D, 11K179D, 07K179D, 06K179D, 06K179D, 03K201D, 06K200D, 01K202D, 08K200D, 09K179D, 06K205D, 07K123D, 08K199D, 09K123D, 09K198D E 10K177D.

Según Agência Nacional de Vigilância Sanitária (ANVISA), este evento ( alerta de seguridad ) involucró a un dispositivo médico en Brazil que fue producido por Abbott Diabetes Care – ADC.

¿Qué es esto?

Las alertas proporcionan información importante y recomendaciones sobre los productos. Aunque se haya emitido una alerta, esto no significa necesariamente que el producto se considera peligroso. Las alertas de seguridad, dirigidas a trabajadores de la salud y a usuarios, pueden incluir retiro de equipos. Pueden ser escritas por los fabricantes, pero también por funcionarios del área de salud.

Más información acerca de la data acá
  • Tipo de evento
    Safety alert
  • ID del evento
    887
  • Fecha
    2007-09-17
  • País del evento
  • Fuente del evento
    ANVISA
  • URL de la fuente del evento
  • Notas / Alertas
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notas adicionales en la data
    Field action initiated because of Abbott Diabetes Care's ADC observation of a higher-than-expected rate of monitors exhibiting damage to the liquid crystal display (LCD) when subjected to a standard drop test. This was observed during a qualification of a new part, where part of the design check drop test involves a standard drop test. As a consequence of this observed fault of display damage, the ADC conducted the same test in the production of monitors and the same display damage was observed. However the monitors affected were only those manufactured after January 31, 2007. At the time the production monitors were identified as being affected, all monitors production and distribution were discontinued and an investigation was initiated.
  • Causa
    A higher-than-expected rate of monitors exhibiting damage to the liquid crystal display (lcd) when subjected to a standard drop test.
  • Acción
    Abbott Diabetes Care - ADC plans to insert a flyer flyer on affected lots of inventory in Brazil and also enable customers to receive the exchange of the product under warranty.

Manufacturer

  • Source
    ANVSANVISA