Alerta De Seguridad para Product name: Tina-quant HbA1c Gen.2 Technical name: HEMOGLOBIN GLICED Registration number ANVISA: Hazard class: II Model affected: Serial numbers affected: Lots distributed in Brazil for Product 4528123190: 11642201; 12661801; 6,017,2801; 60450301; 60801901; 6,089,850; 61710701; 68050801; 69089201; 69127401; 69337401; 69655201; 69986701; 69989301. Lots distributed in Brazil for Product 5401640190: 11641901; 60224601; 60802301; 61061101; 61711001; 69128101; 69337901; 69655601.

Según Agência Nacional de Vigilância Sanitária (ANVISA), este evento ( alerta de seguridad ) involucró a un dispositivo médico en Brazil que fue producido por Roche Diagnóstica Brasil Ltda; Roche Diagnostics GmBH.

¿Qué es esto?

Las alertas proporcionan información importante y recomendaciones sobre los productos. Aunque se haya emitido una alerta, esto no significa necesariamente que el producto se considera peligroso. Las alertas de seguridad, dirigidas a trabajadores de la salud y a usuarios, pueden incluir retiro de equipos. Pueden ser escritas por los fabricantes, pero también por funcionarios del área de salud.

Más información acerca de la data acá
  • Tipo de evento
    Safety alert
  • ID del evento
    2154
  • Fecha
    2016-10-21
  • País del evento
  • Fuente del evento
    ANVISA
  • URL de la fuente del evento
  • Notas / Alertas
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notas adicionales en la data
    Recommendations to the customer / user: - Ensure that adequate storage conditions are maintained as indicated in the instructions for use: 2 - 8 ° C. - Based on this issue, which is occurring sporadically, customers who use the Tina reagent -what HbA1c Gen.2 are now directed to perform QC measurements on each respective bottle of reagent packaging in order to detect potentially affected products. - If the QC can not be kept within the specified range, the cassette of the respective vial should not be used for measurements. - The use of Tina-quant HbA1c Gen.3 reagent in cobas c systems is with appropriate solutions as it is unaffected by this issue. For further questions and support, please do not hesitate to call our CEAC Service Center - 08007720295 of Roche Diagnóstica Brasil Ltda. #### Update of the field action: UPDATED ON 10/27/2017, the company presented the report of completion of the field action, with sending of necessary evidences and corrective actions.
  • Causa
    With this notification, roche diagnostica informs its customers about a possible increase in quality control (qc) recovery and patient samples in single packs of several lots of the tb-quant hba1c gen. 2 reagent when frozen. due to this issue, an update on the instruction instruction will be required for qc measurements on each respective reagent package vial to be performed in order to detect potentially affected products.
  • Acción
    Field Action Code SBN-CPS-2016-019 triggered under the responsibility of the company Roche Diagnóstica Brasil Ltda. Company will update, correct or supplement the instructions for use.

Manufacturer