Alerta De Seguridad para Product name: Vidas Anti HBS Total II (60 T) Technical name: Lives Anti HBS Total II Registration number ANVISA: 1058120684 Hazard Class: IV Affected Model: Not applicable Serial Numbers Affected: 1005262630

Según Agência Nacional de Vigilância Sanitária (ANVISA), este evento ( alerta de seguridad ) involucró a un dispositivo médico en Brazil que fue producido por bioMperieux Brasil Ind e Com de Prod Laboratoriais Ltda.; bioMèrieux S.A..

¿Qué es esto?

Las alertas proporcionan información importante y recomendaciones sobre los productos. Aunque se haya emitido una alerta, esto no significa necesariamente que el producto se considera peligroso. Las alertas de seguridad, dirigidas a trabajadores de la salud y a usuarios, pueden incluir retiro de equipos. Pueden ser escritas por los fabricantes, pero también por funcionarios del área de salud.

Más información acerca de la data acá
  • Tipo de evento
    Safety alert
  • ID del evento
    2321
  • Fecha
    2017-06-27
  • País del evento
  • Fuente del evento
    ANVISA
  • URL de la fuente del evento
  • Notas / Alertas
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notas adicionales en la data
    1- Distribute the information of possible false result to the team members of the laboratories, including other institutions to which they have sent the kits. 2- Return the units that are still in the stock of the customers, units that composed the AWB # 2AWN533, in question, since only these had problems in the transportation to Brazil. 3- Discuss with the Director of the laboratory what decision will be taken regarding the analyzes performed previously with the units received from this lot (which were part of the AWB mentioned above). If you wish to retake the tests contact bioMérieux Brasil. 4- Complete the form to receive the information of recollection and send it back to the Integrated Customer Service of bioMérieux Brazil. ### Update of the field action: UPDATED ON 10/23/2017, the company presented the report of completion of the field action proving the sending of the safety notice to the client with evidence of science and all actions completed.
  • Causa
    An anomaly was detected internally during the evaluation of the data obtained from the dataloggers that accompany the loads of this product. we have identified that the imported product was imported into brazil through awb # 2awn533. this shipment was exposed to excess temperature, having been exposed to temperatures above 41.8 ° c.
  • Acción
    Field Action Code FSCA3509 triggered under the responsibility of the company bioMerieux Brasil Ind e Com de Prod Laboratoriais Ltda. Company will make correction in the field.