Alerta De Seguridad para RAPIDLab 1200 SYSTEMS - Registration No. 10345160455; RAPIDPoint 400 Series - Registration No. 10345160460; RAPIDPoint 500 SYSTEM - Registration No. 10345161877. Serial numbers attached to the alert.

Según Agência Nacional de Vigilância Sanitária (ANVISA), este evento ( alerta de seguridad ) involucró a un dispositivo médico en Brazil que fue producido por Siemens Healthcare Diagnósticos Ltda..

¿Qué es esto?

Las alertas proporcionan información importante y recomendaciones sobre los productos. Aunque se haya emitido una alerta, esto no significa necesariamente que el producto se considera peligroso. Las alertas de seguridad, dirigidas a trabajadores de la salud y a usuarios, pueden incluir retiro de equipos. Pueden ser escritas por los fabricantes, pero también por funcionarios del área de salud.

Más información acerca de la data acá
  • Tipo de evento
    Safety alert
  • ID del evento
    1313
  • Fecha
    2013-10-16
  • País del evento
  • Fuente del evento
    ANVISA
  • URL de la fuente del evento
  • Notas / Alertas
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notas adicionales en la data
    According to the company's statement, mistakes in identifying the sample may lead to the release of incorrect patient results and inappropriate treatment. The clinical results of blood gas analyzers are not interpreted in isolation and inadequate conduct is rare since the physician, through other findings, knows the expected result. Blood gas analyzers are employed in clinical practice and the tests are used to correlate clinical and biochemical findings with optimal treatment and conduct to the patient. Changes in therapy, such as oxygenation rate, would be followed by a repeat test. Patient monitoring would also alert the physician to the potentially discordant outcome, preventing the occurrence of harm.
  • Causa
    Siemens has identified that users may inadvertently enter a wrong patient id on the demographic data editing screen on the rapidpoint® 400, rapidpoint® 500 or rapidlab® 1200 series blood gas analyzers, which could ultimately lead to to an error in the identification of the sample.
  • Acción
    To mitigate misidentification, Siemens directs its customers to remove the Patient List button selection; to check demographic data before leaving the demographic data editing screen, to enable only barcode entry, and to keep the letter with the laboratory records. In addition, a mandatory software update must be performed according to the availability of the software.