Alerta De Seguridad para Tacrolimus Dimension Reagent - Registration nº 10345161332 - Class of Risk II

Según Agência Nacional de Vigilância Sanitária (ANVISA), este evento ( alerta de seguridad ) involucró a un dispositivo médico en Brazil que fue producido por Siemens Healthcare Diagnostics Inc..

¿Qué es esto?

Las alertas proporcionan información importante y recomendaciones sobre los productos. Aunque se haya emitido una alerta, esto no significa necesariamente que el producto se considera peligroso. Las alertas de seguridad, dirigidas a trabajadores de la salud y a usuarios, pueden incluir retiro de equipos. Pueden ser escritas por los fabricantes, pero también por funcionarios del área de salud.

Más información acerca de la data acá
  • Tipo de evento
    Safety alert
  • ID del evento
    1265
  • Fecha
    2013-08-23
  • País del evento
  • Fuente del evento
    ANVISA
  • URL de la fuente del evento
  • Notas / Alertas
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notas adicionales en la data
    Inaccuracy is typically characterized by one or more high TACR results after a period of equipment inactivity. Subsequent results from the same well may show results with decreased values. Falsely decreased or falsely elevated test results may lead to impaired immunosuppressive medication dosage. Increasing the dosage may cause toxicity in the patient's organ. Decreased dosage may lead to organ rejection. If used after long periods of on-board storage (> 2 days) combined with standby periods of the equipment can increase the magnitude of elevation and decrease of results, ie the reduction of on-board stability can result in inaccuracy and in the inaccuracy of quality controls and patient outcomes. Accordingly, imprecision is typically characterized by one or more high TACR results after a period of equipment inactivity. Subsequent results from the same well may show results with decreased values. High discordant results can be raised 2-fold from the expected value and discordant low values ​​can be lowered by fifty percent. The affected lots, that is, the remaining inventory of lots prior to BB4087 must be segregated for collection. As a temporary measure, in order not to interrupt the laboratory routine, lots of reagent cartridge from lot BB4087 will present an "alert card", indicating that the on-board stability is limited to 8 hours. Customers using these batches with the "alert card" should load the Flex reagent cartridges immediately prior to use and remove it after 8 hours. It is also advised not to carry more reagents than the laboratory will use within 8 hours and not pre-hydrate the Tacrolimus Dimension reagents. For further clarification: 0800 129 633. E-mail: subjects. regulatorios.br@siemens.com. Access the warning letter: http://portal.anvisa.gov.br/wps/wcm/connect/ffc77c804ff46d0b8fd3ef6d6e8afaaa/Carta+Portugues+DC+13+08.pdf?MOD=AJPERES
  • Causa
    Results may become inaccurate when the tacrolimus flex reagent is stored on-board in the equipment for 2 days or more. existing and affected lots in brazil: fa3197, fa3267, fb3267, ga3120 - gb3099, gb3176.
  • Acción
    Gathering and destruction. The alert card was adopted as a temporary measure in order not to interrupt the routine of the laboratory while Siemens seeks a definitive solution to this issue.