Alerta De Seguridad para Technical Name: Angiography Equipment. ANVISA registration number: 10345162020. Risk class: III. Affected Model: Artis One. Serial Number Affected: 82190; 82235; 82176; 82062; 82032; 82197; 82187; 82046; 82030; 82136; 82200; 82063; 82090; 82092; 82170; 82216; 82048; 82110; 82143; 82084; 82058; 82067; 82161

Según Agência Nacional de Vigilância Sanitária (ANVISA), este evento ( alerta de seguridad ) involucró a un dispositivo médico en Brazil que fue producido por Siemens Healthcare Diagnósticos Ltda; Siemens Shenzhen Magnetic Ressonance Ltd.(SSMR).

¿Qué es esto?

Las alertas proporcionan información importante y recomendaciones sobre los productos. Aunque se haya emitido una alerta, esto no significa necesariamente que el producto se considera peligroso. Las alertas de seguridad, dirigidas a trabajadores de la salud y a usuarios, pueden incluir retiro de equipos. Pueden ser escritas por los fabricantes, pero también por funcionarios del área de salud.

Más información acerca de la data acá
  • Tipo de evento
    Safety alert
  • ID del evento
    2485
  • Fecha
    2018-02-22
  • País del evento
  • Fuente del evento
    ANVISA
  • URL de la fuente del evento
  • Notas / Alertas
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notas adicionales en la data
    We recommend the distribution of the letter and instruction of the technical team so that they are aware of the problem and in case of doubts should contact the technical support of Siemens If you want to notify technical complaints and adverse events use the channels below: Notivisa: Notices of adverse events ( EA) and technical complaints (QT) for products subject to Sanitary Surveillance should be made through the NOTIVISA System. To access the System, you must register and select the Health Professional option, if you are a liberal professional or the Institution / Entity option, if you are a professional of an institution / entity. Technovigilance System: Patient or citizen can notify through the Technovigilance System / SISTEC access through the link Additional information: -Date of identification of the problem by the company: 05/01/2018 - Date of notification notice for Anvisa: 02/01/2018 The company that owns the affected product is responsible for contacting its customers in a timely manner. to ensure the effectiveness of the Field Action in progress. It stands out the joint responsibility of the distribution chain and use of health products in maintaining its quality, safety and efficacy, as well as the effectiveness of the Field Action, expressed by RDC 23/2012: "(...) Art 2 ° A holder of registration of health product is the holder of the registration / registration of health product with Anvisa. Single paragraph. The registration holder, as well as other agents involved from the production to the use of the product, or discarding it when applicable, are jointly and severally responsible for maintaining the quality, safety and efficacy of the health products to the final consumer. Art. 12 Distributors of health products shall forward to the registration holder, in a timely manner, the distribution map and other information requested for the notification and execution of field actions. (...) "
  • Causa
    The artis upper table top compartment is mounted on the top of the table base and secured with a screw. a locking ring is used to secure the bolt. if the snap ring is installed incorrectly and if the table movement is forced in a non-vertical direction, the table top can slide to the floor after a long period of use. however, for the artis one, there is no table-tilt movement function. therefore the possibility of table movements in a direction other than vertical is remote.
  • Acción
    Field Action Code AX069 / 17 / S & AX071 / 17 / S triggered under the responsibility of Siemens Healthcare Diagnósticos Ltda. Will make field correction

Manufacturer