Alerta De Seguridad para Technical Name: PARAMETERS COMBINED ON THE SAME PRODUCT - CLASS II ANVISA Registration Number: 10345160165 Hazard Class: II Affected Model: Commercial presentations of 2 vials of high calibrator and low calibrator (Calibrator A) Advia Centaur, ACS: 180 Systems 5.0mL / unit) and 6 bottles of high calibrator and low calibrator (5.0mL / unit). Serial numbers affected: 07996A90; 21172A90; 27221A90; 43871A90; 44366A90

Según Agência Nacional de Vigilância Sanitária (ANVISA), este evento ( alerta de seguridad ) involucró a un dispositivo médico en Brazil que fue producido por Siemens Healthcare Diagnósticos S.A; Siemens Healthcare Diagnostics Inc..

¿Qué es esto?

Las alertas proporcionan información importante y recomendaciones sobre los productos. Aunque se haya emitido una alerta, esto no significa necesariamente que el producto se considera peligroso. Las alertas de seguridad, dirigidas a trabajadores de la salud y a usuarios, pueden incluir retiro de equipos. Pueden ser escritas por los fabricantes, pero también por funcionarios del área de salud.

Más información acerca de la data acá
  • Tipo de evento
    Safety alert
  • ID del evento
    2169
  • Fecha
    2017-02-09
  • País del evento
  • Fuente del evento
    ANVISA
  • URL de la fuente del evento
  • Notas / Alertas
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notas adicionales en la data
    Customers using the FT4 assay on the ADVIA Centaur, ADVIA Centaur XP & ADVIA Centaur XPT systems should discontinue use of the field affected lots for the FT4 assay. For the performance of this test, batch calibrators of numbering greater than the affected lot should be used. Customers can continue to use the affected Calibrator A batches for the Free T3 (FT3), Total T3 (T3), Total T4 (T4) and T uptake (Tup) assays for the ADVIA Centaur systems. Customers using the FT4 assay on the ADVIA Centaur CP system can continue to use all available Calibrator A batches. ### UPDATED ON 10/26/2017, the company presented the completion report of the field action proving the sending of the security warning to the client with evidence of science and all actions completed.
  • Causa
    Siemens healthcare diagnostics has confirmed a negative bias for advia centaur free t4 (ft4) when using the a calibrator lots listed above on the advia centaur, advia centaur xp and advia centaur xpt systems. in addition, siemens healthcare diagnostics has confirmed a potential for calibration failures due to the% cvs of the calibrator rlus above the limit when using the calibrator a lots quoted above with the free t4 (ft4) assay. the performance of the free t4 assay (ft4) is not affected when the affected batches of the calibrator are used in the advia centaur cp system.
  • Acción
    Field Action Code CC 17-04 triggered under the responsibility of the company Siemens Healthcare Diagnósticos SA Company will make letter to the client.

Manufacturer