Alerta De Seguridad para Trade name: CONTROL IMMUNODIAGNOSTICO VITROS * FOR HBsAg. ANVISA registry number: 10132590637. Risk class: IV. Product code 6800598. Lots affected and imported in Brazil: 0860, 0870, 0880, 0890, 0908. Lot lot affected: 352 units.

Según Agência Nacional de Vigilância Sanitária (ANVISA), este evento ( alerta de seguridad ) involucró a un dispositivo médico en Brazil que fue producido por JOHNSON & JOHNSON DO BRASIL IND. E COM. DE PROD. PARA SAÚDE LTDA; Ortho Clinical Diagnostics.

¿Qué es esto?

Las alertas proporcionan información importante y recomendaciones sobre los productos. Aunque se haya emitido una alerta, esto no significa necesariamente que el producto se considera peligroso. Las alertas de seguridad, dirigidas a trabajadores de la salud y a usuarios, pueden incluir retiro de equipos. Pueden ser escritas por los fabricantes, pero también por funcionarios del área de salud.

Más información acerca de la data acá
  • Tipo de evento
    Safety alert
  • ID del evento
    2025
  • Fecha
    2016-11-25
  • País del evento
  • Fuente del evento
    ANVISA
  • URL de la fuente del evento
  • Notas / Alertas
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notas adicionales en la data
    According to company information: This is a situation in which the use of the product is not likely to cause adverse health consequences. Action required of laboratories that have the affected lot: • Discontinue use and discard the remaining inventory of affected batches of VITROS HBsAg Controls. (NOTE: Until you receive your replacement order, store the open-reconstituted bottles in the frozen form for up to 4 weeks.) • Forward this notification if you have provided the affected lot (s) outside your laboratory. • Place this notification next to each VITROS System that processes the VITROS Controls for HBsAg • Fill out and return the Receipt Confirmation form. If there is no alternative batch of VITROS HBsAg Controls in your inventory, it is acceptable to continue to use the affected lot (s) by following these instructions: • Control results must be within acceptable limits on the day of reconstitution (Day 1). • Store the open-reconstituted vials in the frozen form for up to 4 weeks. We recommend measuring the aliquot of the open-reconstituted material before freezing, so that several aliquots of the same vial can be used. If you wish to notify technical complaints and adverse events use the following channels: Notivisa: Notices of adverse events (AE) and technical complaints (QT) for products subject to Sanitary Surveillance should be made through the NOTIVISA System. To access the System, you must register and select the Health Professional option, if you are a liberal professional or the Institution / Entity option, if you are a professional of an institution / entity. Technovigilance System: Patient or citizen can notify through the Technovigilance System / SISTEC access through the link
  • Causa
    The company noticed through its global customer complaints system that signal / cutoff (s / c) results increased when negative controls were used. the investigation confirmed that the results generated with the use of these negative control lots showed a deviation in the mean results (s / c) as the product approached its 5-day open-reconstituted limit. the increased results (s / c) may also be observed with the positive control.
  • Acción
    Field Action No. 16000108, triggered under the responsibility of Johnson & Johnson. Field Action Hazard Rating: III Field Action Rating: Field Correction This field action identified internally by the company by the 16000108 encoding results are signal / cutoff (s / c) results increased when used the negative controls. The investigation confirmed that the results generated with the use of these negative control lots showed a deviation in the mean results (s / c) as the product approached its 5-day open-reconstituted limit. The increased results (s / c) may also be observed with the positive control.