Alerta De Seguridad para Trade name: Disposable applicator anchor - Meta Bio ;. ANVISA registration number: 80034760032 ;. Risk class: III. Affected model (s): 2831-020; 2831-027; 2831-030; 2831-035; 2831-040; 2831-045; 2831-050. Lot (s): Distribution map attached.

Según Agência Nacional de Vigilância Sanitária (ANVISA), este evento ( alerta de seguridad ) involucró a un dispositivo médico en Brazil que fue producido por Meta Bio Industrial Ltda..

¿Qué es esto?

Las alertas proporcionan información importante y recomendaciones sobre los productos. Aunque se haya emitido una alerta, esto no significa necesariamente que el producto se considera peligroso. Las alertas de seguridad, dirigidas a trabajadores de la salud y a usuarios, pueden incluir retiro de equipos. Pueden ser escritas por los fabricantes, pero también por funcionarios del área de salud.

Más información acerca de la data acá
  • Tipo de evento
    Safety alert
  • ID del evento
    1779
  • Fecha
    2015-12-23
  • País del evento
  • Fuente del evento
    ANVISA
  • URL de la fuente del evento
  • Notas / Alertas
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notas adicionales en la data
    As registration holder, Disposable Applicator Anchor products manufactured by Meta Bio with the use of thread if sutures in UHMWPE (Ultra High Molecular Weight Polyethylene) material meet all the biocompatibility and mechanical requirements established by the reference standards, as well as to all use / application requirements, according to a comparative evaluation performed on the basis of national and international studies performed by medical researchers. At first, it is not necessary to follow the implanted patients, since the product will not go into failure mode after its implantation, that is, the suture can only rupture at the time of implantation.
  • Causa
    As evidenced by inspection carried out, the products of registration no. 80034760032 (disposable applicator anchor - meta bio) were manufactured using polyethylene suture wires (uhmwpe), this being divergent from the anvisa registry.
  • Acción
    a) Classification of the risk: Class III b) Classification of the field action: collection of the products in the market with action of return to the manufacturer; c) Field action code: Anchor / 2015; d) recommendations to users and patients: payment of the remaining stock balance of the products subject to this notification to our Distributors. At first, it is not necessary to follow the implanted patients, since the product will not go into failure mode after its implantation, that is, the suture can only rupture at the time of implantation.

Manufacturer