Alerta De Seguridad para Trade name: Efficia Vital Signs Monitor. Technical Name: Vital Signs Monitor. ANVISA registration number: 10216710305. Risk class: III. Model affected: CM120, CM150. Serial numbers affected: CM120: CN44701077 to CN62614750. CM150: CN50401316 to CN62609331

Según Agência Nacional de Vigilância Sanitária (ANVISA), este evento ( alerta de seguridad ) involucró a un dispositivo médico en Brazil que fue producido por Philips Medical Systems Ltda; Philips Goldway (Shenzhen) Industrial Inc..

¿Qué es esto?

Las alertas proporcionan información importante y recomendaciones sobre los productos. Aunque se haya emitido una alerta, esto no significa necesariamente que el producto se considera peligroso. Las alertas de seguridad, dirigidas a trabajadores de la salud y a usuarios, pueden incluir retiro de equipos. Pueden ser escritas por los fabricantes, pero también por funcionarios del área de salud.

Más información acerca de la data acá
  • Tipo de evento
    Safety alert
  • ID del evento
    2340
  • Fecha
    2017-07-21
  • País del evento
  • Fuente del evento
    ANVISA
  • URL de la fuente del evento
  • Notas / Alertas
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notas adicionales en la data
    Since the intended use of the monitor is intended for healthcare establishments under the supervision of qualified users (hospital staff), the use of monitors should be discontinued until the software is updated to eliminate the problem. If it is necessary that they be kept in use for a period of time, we recommend that the monitors be monitored constantly to identify if the monitor is faulty to trigger the apnea alarm. Customers who own these monitors will receive an information letter. If you wish to notify technical complaints and adverse events use the following channels: Notivisa: Notices of adverse events (AE) and technical complaints (QT) for products subject to Sanitary Surveillance should be made through the NOTIVISA System. To access the System, you must register and select the Health Professional option, if you are a liberal professional or the Institution / Entity option, if you are a professional of an institution / entity. Technovigilance System: Patient or citizen can notify through the Technovigilance System / SISTEC access through the link - Date of identification of the problem by the company: 02/28/2017 - Date of notification notice to Anvisa: 07/21/2017 The company that owns the affected product is responsible for contacting its customers in a timely manner to ensure the effectiveness of the Field Action in progress. It stands out the joint responsibility of the distribution chain and use of health products in maintaining its quality, safety and efficacy, as well as the effectiveness of the Field Action, expressed by RDC 23/2012: "(...) Art 2 ° A holder of registration of health product is the holder of the registration / registration of health product with Anvisa. Single paragraph. The registration holder, as well as other agents involved from the production to the use of the product, or discarding it when applicable, are jointly and severally responsible for maintaining the quality, safety and efficacy of the health products to the final consumer. Art. 12 Distributors of health products shall forward to the registration holder, in a timely manner, the distribution map and other information requested for the notification and execution of field actions. (...) "
  • Causa
    The reported problem occurs in the efficia vital signs monitor, models cm120 and cm150, software version: b.01.00 (041), when these models are in use with rack efficia (ref 863331) and external etco2 modules from dixtal: capnostat 5 (ref dx-alco2-0) and microstream (ref dx-alco2-1). philips healthtech - pcms - vss brazil has identified, through internal testing, a software anomaly that prevents the sound alarm from being triggered. under certain circumstances, a software error may lead to a situation in which the sound of the apnea alarm is not emitted. the visual alarm will continue to function normally. other related alarms (low fr, low etco2, etc.) are not affected by the problem. to date, there are no reports of adverse events related to this problem.
  • Acción
    Field Action Code DEDF1704 triggered under the responsibility of Philips Medical Systems Ltd. Will update the Software.