Alerta De Seguridad para Trade Name: Harmonic Ace Shears and Adaptive Tissue Technology. Technical Name: Electrosurgical Cable (Power greater than 50W). ANVISA registration number: 80145901455. Risk Class: III. Model affected: HAR23; HAR36. Affected Serial Numbers: Codes and batch numbers related to each code, for individually marketed affected products, and for kits containing affected products are shown in the Customer Letter attached to this Alert.

Según Agência Nacional de Vigilância Sanitária (ANVISA), este evento ( alerta de seguridad ) involucró a un dispositivo médico en Brazil que fue producido por Johnson & Johnson do Brasil Ind. e Com. de Prod. para Saúde Ltda.; Ethicon Endo-Surgery LLC.

¿Qué es esto?

Las alertas proporcionan información importante y recomendaciones sobre los productos. Aunque se haya emitido una alerta, esto no significa necesariamente que el producto se considera peligroso. Las alertas de seguridad, dirigidas a trabajadores de la salud y a usuarios, pueden incluir retiro de equipos. Pueden ser escritas por los fabricantes, pero también por funcionarios del área de salud.

Más información acerca de la data acá
  • Tipo de evento
    Safety alert
  • ID del evento
    2578
  • Fecha
    2018-05-29
  • País del evento
  • Fuente del evento
    ANVISA
  • URL de la fuente del evento
  • Notas / Alertas
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notas adicionales en la data
    The company that holds the registry recommends the following actions: 1. Inspect the inventory immediately to identify the existence of products subject to this recall; 2. After identification, segregate the units and report the problem to the relevant staff of the surgical or material management center; 3. Fill in the Commercial Response Form (FRC) (Annex 2), confirming receipt of the notice within three (3) business days. 4. Keep the communication in a visible place until all batches of products subject to recall have been returned to the company that holds the record. 5. Immediately return all batches of unused and recalled Scissors that are in your inventory until August 31, 2018. If you wish to notify technical complaints and adverse events use the following channels: Notivisa: Notices of adverse events (EA ) and technical complaints (QT) for products subject to Sanitary Surveillance should be made through the NOTIVISA System (http://portal.anvisa.gov.br/notivisa). To access the System, you must register and select the Health Professional option, if you are a liberal professional or the Institution / Entity option, if you are a professional of an institution / entity. Technovigilance System: Patient or citizen can notify through the Technovigilance System / SISTEC access through the link Complementary Information: - Date of identification of the problem by the company: 09/05/2018 - Date of notification notice to Anvisa: 05/28/2018 The company that owns the affected product is responsible for contacting its customers in a timely manner. to ensure the effectiveness of the Field Action in progress. It stands out the joint responsibility of the distribution chain and use of health products in maintaining its quality, safety and efficacy, as well as the effectiveness of the Field Action, expressed by RDC 23/2012: "(...) Art 2 ° A holder of registration of health product is the holder of the registration / registration of health product with Anvisa. Single paragraph. The registration holder, as well as other agents involved from the production to the use of the product, or discarding it when applicable, are jointly and severally responsible for maintaining the quality, safety and efficacy of the health products to the final consumer. Art. 12 Distributors of health products shall forward to the registration holder, in a timely manner, the distribution map and other information requested for the notification and execution of field actions. (...) "
  • Causa
    According to the company holding the registry, there are some devices contained in the aforementioned lots that were assembled with an internal component that can result in continuous or inadvertent activation of the device. informs that such a situation may cause inadvertent mechanical or thermal damage to the unintentional tissue if continuous or inadvertent activation occurs during surgery.
  • Acción
    Field Action Code 1188346 under the responsibility of Johnson & Johnson do Brasil Ind. And Com. De Prod. for Health Ltda. Gathering. Undoing.