Alerta De Seguridad para Trade name: Linear Accelerator Elekta. Technical Name: Linear Accelerator. ANVISA registration number: 80569320007. Risk class: III. Model affected: Elekta Axesse; Infinity; Precise; Synergy; Synergy Platform; Synergy S; Versa HD ;. Serial numbers affected: See attached list.

Según Agência Nacional de Vigilância Sanitária (ANVISA), este evento ( alerta de seguridad ) involucró a un dispositivo médico en Brazil que fue producido por Elekta Medical Systems Comércio e Serviços Radioterapia Ltda; Elekta Ltd.

¿Qué es esto?

Las alertas proporcionan información importante y recomendaciones sobre los productos. Aunque se haya emitido una alerta, esto no significa necesariamente que el producto se considera peligroso. Las alertas de seguridad, dirigidas a trabajadores de la salud y a usuarios, pueden incluir retiro de equipos. Pueden ser escritas por los fabricantes, pero también por funcionarios del área de salud.

Más información acerca de la data acá
  • Tipo de evento
    Safety alert
  • ID del evento
    2428
  • Fecha
    2017-12-12
  • País del evento
  • Fuente del evento
    ANVISA
  • URL de la fuente del evento
  • Notas / Alertas
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notas adicionales en la data
    Read and follow the Important Field Security Notice (200-01-103-082) and be aware of the matter, and discuss it by completing and signing the document of designation with your Elekta representative as soon as possible. Allow Elekta Service representatives to access the machine to install the modification as soon as the solution is available. If you want to notify technical complaints and adverse events use the following channels: Notivisa: Notices of adverse events (AEs) and technical complaints (QT) for products subject to Sanitary Surveillance should be done through the NOTIVISA System. To access the System, you must register and select the Health Professional option, if you are a liberal professional or the Institution / Entity option, if you are a professional of an institution / entity. Technovigilance System: Patient or citizen can notify through the Technovigilance System / SISTEC access through the link Additional information: - Date of identification of the problem by the company: 10/26/2017 - Date of notification notice to Anvisa: 12/12/2017 The company that owns the affected product is responsible for contacting its customers in a timely manner. to ensure the effectiveness of the Field Action in progress. It stands out the joint responsibility of the distribution chain and use of health products in maintaining its quality, safety and efficacy, as well as the effectiveness of the Field Action, expressed by RDC 23/2012: "(...) Art 2 ° A holder of registration of health product is the holder of the registration / registration of health product with Anvisa. Single paragraph. The registration holder, as well as other agents involved from the production to the use of the product, or discarding it when applicable, are jointly and severally responsible for maintaining the quality, safety and efficacy of the health products to the final consumer. Art. 12 Distributors of health products shall forward to the registration holder, in a timely manner, the distribution map and other information requested for the notification and execution of field actions. (...) "
  • Causa
    Potential for iviewgt ™ and xvi detector arms to extend due to gravity elekta has identified that if the microswitch inside the pcb assembly, the middle arm's solenoid switch is configured incorrectly, there is a potential for an extension control of the iviewgt ™ / xvi detector arms. when the detector arm is fully retracted, there is no risk.
  • Acción
    Field Action Code 3 triggered under the responsibility of Elekta Medical Systems Comércio e Serviços Radiotherapy Ltda. Will make field correction

Manufacturer