Alerta De Seguridad para Trade name: VENTILATION SYSTEM. ANVISA registration number: 80259110072. Model affected: Servo-i. . Commercial Name: Servo-air Ventilation System. ANVISA registration number: 80259110124. Affected model: Servo-Air. . Trade name: SERVO-I PULMONARY VENTILATOR. ANVISA registration number: 80259110006. Model affected: Servo-s. . Trade name: VENTILADOR PULMONAR SERVO-S. ANVISA registration number: 80259110034. Affected model: Servo-u / n. . Technical Name: Pressure and Volume Fan. Risk class: III. Serial number affected: See distribution map

Según Agência Nacional de Vigilância Sanitária (ANVISA), este evento ( alerta de seguridad ) involucró a un dispositivo médico en Brazil que fue producido por Maquet do Brasil Equipamentos Medicos LTDA.; Maquet Critical Care AB.

¿Qué es esto?

Las alertas proporcionan información importante y recomendaciones sobre los productos. Aunque se haya emitido una alerta, esto no significa necesariamente que el producto se considera peligroso. Las alertas de seguridad, dirigidas a trabajadores de la salud y a usuarios, pueden incluir retiro de equipos. Pueden ser escritas por los fabricantes, pero también por funcionarios del área de salud.

Más información acerca de la data acá
  • Tipo de evento
    Safety alert
  • ID del evento
    2524
  • Fecha
    2018-03-16
  • País del evento
  • Fuente del evento
    ANVISA
  • URL de la fuente del evento
  • Notas / Alertas
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notas adicionales en la data
    The Servo fan can be used according to the instructions for use, with extra attention to the following: • Never leave the patient unattended when connected to the fan system. • Make sure a resuscitator is readily available. • Act as instructed in the alarms: In case of technical error TE: 11, replace the fan immediately and contact your Getinge / Maquet representative. If you wish to notify technical complaints and adverse events use the following channels: Notivisa: Notices of adverse events (AE) and technical complaints (QT) for products subject to Sanitary Surveillance should be made through the NOTIVISA System. To access the System, you must register and select the Health Professional option, if you are a liberal professional or the Institution / Entity option, if you are a professional of an institution / entity. Technovigilance System: Patient or citizen can notify through the Technovigilance System / SISTEC access through the link Additional information: - Date of identification of the problem by the company: 02/28/2018 - Date of notification notice to Anvisa: 03/16/2018 The company that owns the affected product is responsible for contacting its customers in a timely manner. to ensure the effectiveness of the Field Action in progress. It stands out the joint responsibility of the distribution chain and use of health products in maintaining its quality, safety and efficacy, as well as the effectiveness of the Field Action, expressed by RDC 23/2012: "(...) Art 2 ° A holder of registration of health product is the holder of the registration / registration of health product with Anvisa. Single paragraph. The registration holder, as well as other agents involved from the production to the use of the product, or discarding it when applicable, are jointly and severally responsible for maintaining the quality, safety and efficacy of the health products to the final consumer. Art. 12 Distributors of health products shall forward to the registration holder, in a timely manner, the distribution map and other information requested for the notification and execution of field actions. (...) "
  • Causa
    The registrant has identified that, under certain conditions, an internal communication error may prevent the automatic closing of the safety valve. as a consequence, the ventilation will stop and a high priority alarm will be triggered as well as clinical alarms.
  • Acción
    Field Action Code MCC / 18/001 / IU triggered under the responsibility of Maquet do Brasil Equipamentos Medicos LTDA.

Manufacturer

  • Empresa matriz del fabricante (2017)
  • Source
    ANVSANVISA