Alerta De Seguridad para Trademark: IntelliVue iX Information Center. Technical Name: Software. ANVISA registration number: 10216710250. Risk class: III. Model Affected: M3290B. Affected serial numbers: JMX2031Y0F3 and JMX2039Y1GG.

Según Agência Nacional de Vigilância Sanitária (ANVISA), este evento ( alerta de seguridad ) involucró a un dispositivo médico en Brazil que fue producido por Philips Medical Systems Ltda..

¿Qué es esto?

Las alertas proporcionan información importante y recomendaciones sobre los productos. Aunque se haya emitido una alerta, esto no significa necesariamente que el producto se considera peligroso. Las alertas de seguridad, dirigidas a trabajadores de la salud y a usuarios, pueden incluir retiro de equipos. Pueden ser escritas por los fabricantes, pero también por funcionarios del área de salud.

Más información acerca de la data acá
  • Tipo de evento
    Safety alert
  • ID del evento
    2392
  • Fecha
    2017-10-09
  • País del evento
  • Fuente del evento
    ANVISA
  • URL de la fuente del evento
  • Notas / Alertas
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notas adicionales en la data
    To verify that your Cisco Adaptive Security Appliance (ASA) has been affected by this problem, examine the Cisco firewall serial number. Any device with a serial number below JMX2050000 is affected. There are two ways to get the serial number: the command line interface (CLI) and the visual examination of the chassis. No further action is required on the part of users and patients. If you wish to notify technical complaints and adverse events use the following channels: Notivisa: Notices of adverse events (AE) and technical complaints (QT) for products subject to Sanitary Surveillance should be made through the NOTIVISA System. To access the System, you must register and select the Health Professional option, if you are a liberal professional or the Institution / Entity option, if you are a professional of an institution / entity. Technovigilance System: Patient or citizen can notify through the Technovigilance System / SISTEC access through the link Complementary Information: - Date of identification of the problem by the company: 09/29/2017 - Date of notification notice for Anvisa: 09/10/2017 The company that owns the affected product is responsible for contacting its customers in a timely manner. to ensure the effectiveness of the Field Action in progress. It stands out the joint responsibility of the distribution chain and use of health products in maintaining its quality, safety and efficacy, as well as the effectiveness of the Field Action, expressed by RDC 23/2012: "(...) Art 2 ° A holder of registration of health product is the holder of the registration / registration of health product with Anvisa. Single paragraph. The registration holder, as well as other agents involved from the production to the use of the product, or discarding it when applicable, are jointly and severally responsible for maintaining the quality, safety and efficacy of the health products to the final consumer. Art. 12 Distributors of health products shall forward to the registration holder, in a timely manner, the distribution map and other information requested for the notification and execution of field actions. (...) "
  • Causa
    The defect involves the clock signal component inside the firewall. this component has a high probability of failing on devices that have been running for more than 18 months. if the clock signal component fails, the firewall will no longer function, will not initialize, and will not be able to recover from component failure. this failure will result in a loss of communication between the devices that are separated by the firewall, which can cause the information center to restart.
  • Acción
    Field Action FCO86201775 Code triggered under the responsibility of Philips Medical Systems Ltda. Will perform field correction.

Manufacturer

  • Empresa matriz del fabricante (2017)
  • Source
    ANVSANVISA