Alerta De Seguridad para TROPONIN I TURBO IMMULITE / IMMULITE 1000, Kit for 100 tests, Registration 10345160677, lots 0319; 0321; 0322 and 0323.

Según Agência Nacional de Vigilância Sanitária (ANVISA), este evento ( alerta de seguridad ) involucró a un dispositivo médico en Brazil que fue producido por Siemens Healthcare Diagnósticos Ltda.; Siemens Healthcare Diagnostics Products Ltd.

¿Qué es esto?

Las alertas proporcionan información importante y recomendaciones sobre los productos. Aunque se haya emitido una alerta, esto no significa necesariamente que el producto se considera peligroso. Las alertas de seguridad, dirigidas a trabajadores de la salud y a usuarios, pueden incluir retiro de equipos. Pueden ser escritas por los fabricantes, pero también por funcionarios del área de salud.

Más información acerca de la data acá
  • Tipo de evento
    Safety alert
  • ID del evento
    1482
  • Fecha
    2014-11-26
  • País del evento
  • Fuente del evento
    ANVISA
  • URL de la fuente del evento
  • Notas / Alertas
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notas adicionales en la data
    The overseas manufacturer performed the problem analysis, evaluating all the inherent risks that generated the corrective action in the field. Based on the investigation and mitigation of the risk conducted, this action was classified as a medium risk to health, its severity being moderate and its probability of occurrence probable. The use of the batches of affected reagents may lead to an increase in the frequency of falsely elevated troponin values. Quality control is unlikely to detect this problem. A falsely elevated troponin value may lead to additional treatment or unnecessary diagnostic procedure. Siemens is recommending a review for any troponin sample that is in the sample stability period, and with troponin values ​​higher than the cut-off established by the laboratory, or greater than the 99th percentile, and which has not been confirmed by another method.
  • Causa
    Siemens confirmed that troponin i turbo immulite / immulite 1000 reagents, batches 319, 321, 322, and 323 demonstrated an increase in the frequency of 2-7% in the number of falsely elevated troponin values ​​in patient samples above the 99th percentile > 0.30 ng / ml (μg / l) as published in the instructions for use. quality control is unlikely to detect this problem.
  • Acción
    The company advises customers to discontinue use of the product, segregate any remaining stock from the affected lot of Troponin in order to return these products to Siemens and keep the letter with the laboratory records. In addition, Siemens recommends retaining the letter with the laboratory records and forwarding the notification to anyone who may have access to the products.

Manufacturer