Retiro De Equipo (Recall) de DIRECT DRIVE DISPOSABLE GRASPER

Según Health Canada (via FOI), este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Canada que fue producido por SOUTHMEDIC INC..

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    72546
  • Clase de Riesgo del Evento
    II
  • Fecha de inicio del evento
    2011-07-18
  • País del evento
  • Fuente del evento
    HC
  • Notas / Alertas
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Notas adicionales en la data
  • Causa
    Possible discrepancies in the in the material drying time and consequent injection molding. parameters during the manufacture of the respective locking trigger lots. as a result of these non-conformities there is potential.For the locking trigger to fracture during usage and allow the grasper to remain in a closed position. all units/kits that were built with the suspect lots of triggers have either been put in quarantine or are included in the recall.

Device

  • Modelo / Serial
    Model Catalog: C4130 (Lot serial: 1137511); Model Catalog: C4130 (Lot serial: 1127581); Model Catalog: C4130 (Lot serial: 1126120)
  • Descripción del producto
    Direct Drive Grasper
  • Manufacturer

Manufacturer