Retiro De Equipo (Recall) de FLEXOR CHECK-FLO PERFORMER INTRODUCER SET

Según Health Canada (via FOI), este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Canada que fue producido por COOK (CANADA) INC..

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    22967
  • Clase de Riesgo del Evento
    II
  • Fecha de inicio del evento
    2015-06-09
  • País del evento
  • Fuente del evento
    HC
  • Notas / Alertas
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Notas adicionales en la data
  • Causa
    This recall is being initiated by cook medical due to reports received indicating the introducing dilator component extends beyond the introducer sheath further than intended and does not conform to the specifications. potential adverse events that may occur as a result of the excess dilator length include difficulty in accessing the hepatic vein requiring additional manipulation which could result in vessel injury.

Device

  • Modelo / Serial
    Model Catalog: KCFW-10.0-35-RB-RTPS-100 (Lot serial: 5805365); Model Catalog: KCFW-10.0-35-RB-RTPS-100 (Lot serial: 5770328); Model Catalog: KCFW-10.0-35-RB-RTPS-100 (Lot serial: 5770364)
  • Descripción del producto
    Flexor RTPS Guiding Sheath
  • Manufacturer

Manufacturer