Retiro De Equipo (Recall) de INFINITY ACUTE CARE SYSTEM WORKSTATION NEONATAL CARE - BABYLOG VN500

Según Health Canada (via FOI), este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Canada que fue producido por DRAEGERWERK AG & CO. KGAA.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    15486
  • Clase de Riesgo del Evento
    II
  • Fecha de inicio del evento
    2015-07-08
  • País del evento
  • Fuente del evento
    HC
  • Notas / Alertas
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Notas adicionales en la data
  • Causa
    Draeger became aware about an event in which the battery capacity of the optional ps500 of an iacs workstation critical care (evita infinity v500) did not last as long as expected. the batteries installed in the ps500 depleted much earlier than expected although the battery indicator showed a sufficiently charged battery. the sequential alarms "battery low" followed by "batter depleated" were given however the remaining time until the ventilator shut down was not foreseeable. when the battery depleted totally the power fail alarm was generated. draeger issued a measure in february 2014 and july 2014 regarding battery capacity failures. the root cause for the reduced battery capacity at that time was due to an incorrect charging algorithm that subsequently caused sulfation within the battery. the new battery that was exchanged with that corrective action was tested and qualified. however draeger has learned from the supplier that the batteries of this lot may be affected by a manufacturing defect that may lead to shorter battery life times. as a consequence draeger has decided to replace all batteries for the ps500 again.

Device

  • Modelo / Serial
    Model Catalog: 8415000 (Lot serial: >10 NUMBERS CONTACT MFG)
  • Descripción del producto
    INFINITY ACUTE CARE SYSTEM WORKSTATION NEONATAL CARE - BABYLOG VN500