Retiro De Equipo (Recall) de PROTEUS XR/A GENERAL PURPOSE RADIOGRAPHIC SYSTEM

Según Health Canada (via FOI), este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Canada que fue producido por GENERAL ELECTRIC CANADA (OPERATING AS GE HEALTHCARE).

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    57242
  • Clase de Riesgo del Evento
    II
  • Fecha de inicio del evento
    2014-06-10
  • País del evento
  • Fuente del evento
    HC
  • Notas / Alertas
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Notas adicionales en la data
  • Causa
    If a device is improperly re-installed the collimator may not be properly seated and may loosen and fall. such an event may occur immediately following re-installation during a service maintenance activity but may also occur a few months after a service activity and continued use. a fall of a collimator while the system is in use could result in an injury to a patient or operator. there was a reported incident of a collimator fall which caused a serious patient injury.

Device

  • Modelo / Serial
    Model Catalog: 2259988 (Lot serial: > 10 numbers contact mfg.); Model Catalog: GTC14064PC-2 (Lot serial: > 10 numbers contact mfg.)
  • Descripción del producto
    PROTEUS XR/A GENERAL PURPOSE RADIOGRAPHIC SYSTEM