Retiro De Equipo (Recall) de VITROS CHEMISTRY PRODUCTS 51 FS CHEMISTRY SYSTEM - INSTRUMENT

Según Health Canada (via FOI), este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Canada que fue producido por ORTHO-CLINICAL DIAGNOSTICS.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    78223
  • Clase de Riesgo del Evento
    III
  • Fecha de inicio del evento
    2013-02-19
  • País del evento
  • Fuente del evento
    HC
  • Notas / Alertas
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Notas adicionales en la data
  • Causa
    Ocd determined that extension springs on some vitros systems may be out of the manufacturing specifications. extension springs are part of the slide alignment guide assembly. during an internal evaluation in our manufacturing facility it was discovered that an extension spring had insufficient force. if the extension spring does not provide sufficient force microslides may not be properly positioned within the microslide incubator which has the potential to cause imprecise results. extension springs are just one component in the slide alignment assembly. our evaluation determined that the probability of an out of specification extension spring alone adversely affecting results is unlikely. to date ocd has not received any customer complaints related to this issue.

Device

  • Modelo / Serial
    Model Catalog: 6801375 (Lot serial: > 10 contact manufacturer); Model Catalog: 680 2445 (Lot serial: > 10 contact manufacturer); Model Catalog: 680 2413 (Lot serial: > 10 contact manufacturer)
  • Descripción del producto
    VITROS 51 Chemistry System