Alerta De Seguridad para HeartWare Ventricular Assist System, Controller

Según Department of Health, este evento ( alerta de seguridad ) involucró a un dispositivo médico en Hong Kong que fue producido por HeartWare.

¿Qué es esto?

Las alertas proporcionan información importante y recomendaciones sobre los productos. Aunque se haya emitido una alerta, esto no significa necesariamente que el producto se considera peligroso. Las alertas de seguridad, dirigidas a trabajadores de la salud y a usuarios, pueden incluir retiro de equipos. Pueden ser escritas por los fabricantes, pero también por funcionarios del área de salud.

Más información acerca de la data acá
  • Tipo de evento
    Safety alert
  • Fecha
    2013-05-20
  • Fecha de publicación del evento
    2013-05-20
  • País del evento
  • Fuente del evento
    DH
  • URL de la fuente del evento
  • Notas / Alertas
    Hong Kong data is current through September 2018. All of the data comes from the Department of Health (Hong Kong), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Hong Kong.
  • Notas adicionales en la data
    Medical Device Safety Alert
  • Causa
    Medical device safety alert: heartware ventricular assist system, controller the australia therapeutic goods administration (tga) has posted a medical device safety alert concerning heartware controller, component of a ventricular assist device. the affected product codes are 1400 and 1401au. all serial numbers are affected. heartware has advised of a recent incident involving a patient death thirty-four months post heartware ventricular assist system implantation. the patient was found alone and unresponsive with the controller alarming approximately three hours post event. log file analysis confirmed that the pump had stopped. functional testing of the returned controller showed that the device (including motor control circuits) performed all functions as intended within specification and with no fault alarms or errors. the exact cause of the event cannot be conclusively determined. however, heartware suspects that an electrostatic discharge through the exposed controller power ports during battery replacement caused or contributed to data corruption in the pump motor controller resulting in a loss of commutation in which the motor control software was no longer driving the pump’s motor controller circuit and leading to a pump stop. physicians are reminded of the following: be aware of electrostatic discharge and its potential to cause disruptive and possibly fatal faults to susceptible patients. avoid devices and conditions that may induce strong static discharges (for example, television or computer monitor screens) as electrostatic discharges can damage the electrical parts of the system and cause the left ventricular assist device to perform improperly or stop. always have a backup controller handy and a care giver nearby when changing power sources or controllers. be watchful for unusual changes in power or flow alarms for a period of time following equipment change. according to the local supplier, the affected products were not distributed in hong kong. for details, please visit the following tga website: http://www.Tga.Gov.Au/sara/arn-detail.Aspx?k=rc-2013-rn-00438-1 if you are in possession of the affected products, please contact your supplier for necessary actions. posted on 20 may 2013.

Device

  • Modelo / Serial
  • Descripción del producto
    Medical Device Safety Alert: HeartWare Ventricular Assist System, Controller
  • Manufacturer

Manufacturer

  • Empresa matriz del fabricante (2017)
  • Comentario del fabricante
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.
  • Source
    DH