Alerta De Seguridad para TheraSphere Yttrium 90 Glass Microspheres

Según Department of Health, este evento ( alerta de seguridad ) involucró a un dispositivo médico en Hong Kong que fue producido por Biocompatibles UK.

¿Qué es esto?

Las alertas proporcionan información importante y recomendaciones sobre los productos. Aunque se haya emitido una alerta, esto no significa necesariamente que el producto se considera peligroso. Las alertas de seguridad, dirigidas a trabajadores de la salud y a usuarios, pueden incluir retiro de equipos. Pueden ser escritas por los fabricantes, pero también por funcionarios del área de salud.

Más información acerca de la data acá
  • Tipo de evento
    Safety alert
  • Fecha
    2015-09-23
  • Fecha de publicación del evento
    2015-09-23
  • País del evento
  • Fuente del evento
    DH
  • URL de la fuente del evento
  • Notas / Alertas
    Hong Kong data is current through September 2018. All of the data comes from the Department of Health (Hong Kong), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Hong Kong.
  • Notas adicionales en la data
    Medical Device Safety Alert
  • Causa
    Medical device safety alert: biocompatibles uk therasphere yttrium 90 glass microspheres medical device manufacturer, biocompatibles uk (btg), has issued a medical device safety alert concerning its therasphere yttrium 90 (y-90) glass microspheres. the affected products are identified as follows:- lot number: 1599211; manufacturing date: 03-sep-2015; calibration date: 06-sep-2015; expiry date: 18-sep-2015 the manufacturer has recently become aware, via a customer complaint, of the mislabeling of one dose vial of therasphere y-90 glass microspheres lot 1599211 that was distributed in hong kong. the therasphere vial was labeled 10 gbq and the actual dose was 5 gbq (total radioactivity was incorrect). mislabelling of the therasphere y-90 glass microspheres may result in under- or over-dosing of the patient resulting in the patient not receiving their prescribed dose. under-dosing could result in lack of efficacy, and over-dosing could result in non-target tissue radiation. the manufacturer advises users to verify the accuracy of the dose measurement prior to administration. this recommendation is consistent with the therasphere training and reference material that the manufacturer provides to each user. according to the local supplier, the affected products are distributed in hong kong. if you are in possession of the affected products, please contact your supplier for necessary actions. posted on 23 september 2015.

Device

  • Modelo / Serial
  • Descripción del producto
    Medical Device Safety Alert: Biocompatibles UK TheraSphere Yttrium 90 Glass Microspheres
  • Manufacturer

Manufacturer

  • Empresa matriz del fabricante (2017)
  • Source
    DH