Las notificaciones de seguridad de campo son comunicaciones sobre acciones que se pueden estar tomando en relación con un producto que se encuentra en el mercado, enviadas por los fabricantes de dispositivos médicos o sus representantes. Estas notificaciones son principalmente para trabajadores de la salud, pero también para usuarios. Pueden incluir alertas y retiro de equipos.

Más información acerca de la data acá
  • Tipo de evento
    Field Safety Notice
  • Número del evento
    SN2003(01)
  • Fecha
    2003-02-13
  • País del evento
  • Fuente del evento
    HPRA
  • URL de la fuente del evento
  • Notas / Alertas
    "Irish data is current through April 2019. All of the data comes from the Health Products Regulatory Authority (Ireland), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Ireland."
  • Notas adicionales en la data
  • Causa
    The imb has been advised of several incidents with patients on long-term total parenteral nutrition using a broviac paediatric catheter.  defects developed at a similar point on the catheter, where the hub separated from the catheter.
  • Acción
    The IMB advise that users of this product:  Ensure that all staff dealing with broviac lines are advised of the incidents that have arisen with this device.   Advise all staff and family members dealing with the device to carry out regular inspections of the device.   Ensure that all staff and family members are trained in the use of the catheter. Ensure the device is clamped at the points indicated by the manufacturer in their literature.   Report any indication of problems with the device to the Irish Medicines Board