Retiro De Equipo (Recall) de Abbott Alkaline Wash Solution

Según New Zealand Medicines and Medical Devices Safety Authority, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en New Zealand que fue producido por Abbott Laboratories Inc.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    18527
  • Fecha de inicio del evento
    2015-05-14
  • País del evento
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Abbott Diagnostics Division, Ground Floor, Building D, 4 Pacific Rise, Mt Wellington, AUCKLAND 1640
  • Causa
    Abbott diagnostics is recalling alkaline wash lot number 49059un14 due to leaking product and/or material received with loose bottle caps that may have leaked during shipment. alkaline solution is corrosive and the operator can be exposed when the bottle is leaking.
  • Acción
    Product to be destroyed

Device

  • Modelo / Serial
    Model: List Number: 9D31-20, Affected: Lot Number: 49059UN14
  • Manufacturer

Manufacturer

  • Empresa matriz del fabricante (2017)
  • Comentario del fabricante
    “We are in constant communication with regulatory agencies and competent authorities worldwide which allows us to implement global recalls or in-country communication quickly and effectively,” Abbott, which now owns St. Jude Medical told ICIJ in a statement. In addition to sending global notices to physicians worldwide, we also make sure that product advisories are available online and classification of product recalls and product advisories are determined by global regulatory bodies which can impact the timing in any given country. MD companies follow varying regulations in different countries. In come countries software is not regulated so a recall in one country related to software would not be classified as a recall or field action in another. In addition, review cycles within the regulatory process can be different in each country which can impact communication and recall timing.
  • Source
    NZMMDSA