Retiro De Equipo (Recall) de Agfa Computed Radiography NX Workstation

Según New Zealand Medicines and Medical Devices Safety Authority, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en New Zealand que fue producido por Agfa-Gevaert NV.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    21092
  • Fecha de inicio del evento
    2017-02-14
  • País del evento
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Cass Medical Limited, 803 Paparata Road, Bombay, SOUTH AUCKLAND
  • Causa
    Due to a memory issue a warning message about unexpected error and possible incorrect data integrity is shown to the user. a reboot of the workstation computer after having received the warning message prevents the issue from occurring. in one instance the warning instructions were not followed and resulted in the image of one patient being archived in another patient's record.
  • Acción
    Software to be upgraded

Device

  • Modelo / Serial
    Model: , Affected: Software 3.0.8950
  • Manufacturer

Manufacturer