Retiro De Equipo (Recall) de Boston Scientific AUTOGEN ICD DR Defibrillator, Model D177

Según New Zealand Medicines and Medical Devices Safety Authority, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en New Zealand que fue producido por Boston Scientific Corporation.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    17721
  • Fecha de inicio del evento
    2014-11-19
  • País del evento
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Boston Scientific NZ Ltd, Level 1, 3 Diamond Street, Newton, Auckland
  • Causa
    These devices include the option of enabling a right ventricular automatic threshold (rvat) test to determine the rv pacing threshold and adjust amplitude in an ambulatory setting. if the rvat test feature is enabled and noise signals are continuously sensed within a brief rv noise window following an atrial pace, a patient may not receive effective pacing support until the rvat test ends (i.E., up to 20 cardiac cycles). although no patients have been harmed in the cases reported to date, brief periods of dizziness were reported in one case.
  • Acción
    Software to be upgraded

Device

Manufacturer

  • Empresa matriz del fabricante (2017)
  • Comentario del fabricante
    “We take a patient-first approach to assessing the applicability of every recall and communicate to regulatory bodies in all geographies where the recalled device is sold,” Boston Scientific said in a statement to ICIJ. “We have coordinated several recalls across many countries in a timely manner,” the company said, adding that it complies with all national laws, which can often vary and require different processes for reporting information or taking action on recalls. The company said it uses a rigorous and uniform process to take action on recalls and that “when we initiate a field action (e.g. recall, safety alert), every customer who has received an affected product receives a communication that includes a letter for the physician.”
  • Source
    NZMMDSA