Retiro De Equipo (Recall) de Carestream DRX Revolution X-Ray System

Según New Zealand Medicines and Medical Devices Safety Authority, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en New Zealand que fue producido por Carestream Health Inc.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    17828
  • Fecha de inicio del evento
    2014-12-04
  • País del evento
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Carestream Health New Zealand Ltd, c/- Quigg Partners, Level 7, 28 Brandon St, Wellington
  • Causa
    The manufacturer has received a report that an overseas service engineer observed a loose boom arm during routine maintenance on a drx revolution mobile x-ray system as one of three bolts that secure the boom to the carriage had fractured and the other two bolts were loose causing the boom arm to become loose. no other reports of this type have been received since the device was introduced into commercial distribution in 2009.
  • Acción
    Product to be modified

Device

  • Modelo / Serial
    Model: , Affected: Manufactured prior to 15-Oct-2013