Retiro De Equipo (Recall) de Cook Test Tube Heater, power supply unit

Según New Zealand Medicines and Medical Devices Safety Authority, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en New Zealand que fue producido por William A Cook Australia Pty Ltd.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    17248
  • Fecha de inicio del evento
    2014-09-02
  • País del evento
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Obex Medical Ltd, 303 Manukau Road, Epsom, AUCKLAND
  • Causa
    The power supply of the devices are being replaced because recent testing of the combination of the test tube heater and power supply was found to exceed the prescribed electromagnetic compatibility (emc) limits for radiated emissions., as the level of excess emissions are only slightly over the prescribed limits, and because the exposure to the radio frequency (rf) fields is far below internationally accepted safety levels, it is unlikely that there are any health risks to the user or patient posed by the non-conforming combination of the test tube heater (k-fth-1012) and power supply (k-aps-300). it is also unlikely for the power supply to affect the operation of the test tube heater as this device is designed to have rf immunity greater than the radiation emitted by the power supply.
  • Acción
    Product to be exchanged

Device

Manufacturer