Retiro De Equipo (Recall) de Cordis Empira NC Rx PTCA Catheter

Según New Zealand Medicines and Medical Devices Safety Authority, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en New Zealand que fue producido por Creganna-Tactx Medical (USA).

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    18056
  • Fecha de inicio del evento
    2015-02-09
  • País del evento
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Johnson & Johnson Medical New Zealand Ltd, 507 Mt Wellington Highway, Mt Wellington, AUCKLAND 1060
  • Causa
    Through the investigation of the complaints, cordis and creganna-tactx have observed that the balloon of the affected product, which is covered with a hydrophilic coating, may exhibit resistance to the removal of the balloon cover. weakening of the affected product in the region of the proximal balloon seal may occur when excessive force is applied to remove the balloon cover. use of a weakened catheter may lead to balloon burst, shaft separation or deflation difficulties, which could result in procedural delay, vessel trauma or surgery. there is no concern for patients who have already been treated successfully with the affected product.
  • Acción
    Instructions for use to be updated