Retiro De Equipo (Recall) de Datex GE Carestation Anaesthesia Work Stations

Según New Zealand Medicines and Medical Devices Safety Authority, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en New Zealand que fue producido por Datex-Ohmeda Inc.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    22074
  • Fecha de inicio del evento
    2017-10-12
  • País del evento
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: GE Healthcare Limited, 8 Tangihua Street, Auckland 1010
  • Causa
    It has been identified that an incomplete seal can exist between the co2 absorber and the breathing circuit co2 bypass port assembly of the carestation 600 series systems. an incomplete seal can allow rebreathing of patient gases that have bypassed the co2 absorbent material and could result in unintended elevated levels of inspired co2 (fico2), which could lead to hypercarbia. there have been no injuries reported as a result of this issue., manufacturer is releasing revised parts that minimize the likelihood of incomplete gas flow through the co2 absorbent canister. gehc will correct all affected products by installing updated parts.
  • Acción
    Product to be modified

Device

  • Modelo / Serial
    Model: 630, 650 and 650c, Affected:
  • Manufacturer