Retiro De Equipo (Recall) de Draeger Infinity V500 with optional PS500

Según New Zealand Medicines and Medical Devices Safety Authority, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en New Zealand que fue producido por Draegerwerk AG & Co KGaA.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    19463
  • Fecha de inicio del evento
    2015-11-24
  • País del evento
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Draeger New Zealand Ltd, Unit 4, 24 Bishop Dunn Place, East Tamaki, Auckland 2013
  • Causa
    Draeger has become aware of cases in which the battery capacity of the optional ps500 of an iacs workstation critical care (evita infinity v500, evita v300 and babylog vn500 ventilators) was unexpectedly shortened. the batteries were discharged early, even though the device was indicating an adequate battery state of charge. the devices generate a "battery low" alarm, followed by "battery flat". in exceptional cases, the "battery flat" alarm might be displayed too late, or not at all. after complete discharge of the batteries, the device correctly alerts to failure of the power supply by the auxiliary alarm (piezo, power failure alarm) in all cases., current analysis of the battery data available indicates that an operating time of at least 20 minutes can normally be maintained after the device is disconnected from the power supply, which sufficiently bridges the gap in the event of a mains power failure until the emergency power supply can be activated.
  • Acción
    Software to be upgraded

Device

Manufacturer