Retiro De Equipo (Recall) de Drager D-Vapor

Según New Zealand Medicines and Medical Devices Safety Authority, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en New Zealand que fue producido por Draegerwerk AG & Co KGaA.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    20020
  • Fecha de inicio del evento
    2016-04-19
  • País del evento
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Draeger New Zealand Ltd, Unit 4, 24 Bishop Dunn Place, East Tamaki, Auckland 2013
  • Causa
    As part of our continued market and product monitoring we have become aware of isolated cases in which the unwanted release of evaporating agent was observed by users at the end of the filling procedure. the release may lead to spraying of aerosol into face and eyes of the operator. eye contact with volatile anaesthetics can result in serious irritation; immediate flushing of eyes and consultation of a physician is recommended. recently updated material safety data sheets (msds) of desflurane manufacturers recommend the wearing of safety glasses when handling these substances.
  • Acción
    Instructions for use to be updated