Retiro De Equipo (Recall) de Elekta Integrity System

Según New Zealand Medicines and Medical Devices Safety Authority, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en New Zealand que fue producido por Elekta Limited (England).

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    14511
  • Fecha de inicio del evento
    2013-04-23
  • País del evento
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Elekta Pty Limited - New Zealand Branch, Level 2, Windsor Court , 128-136 Parnell Road, Parnell, AUCKLAND 1052
  • Causa
    During beam delivery, the tolerance values are used to do a check for unwanted movements. these values are known as "static tolerances" which are related to all the digital accelerator geometric movement axes except table column. the problem is that the "static tolerances" from the calibration files (loaded database) are used for beam delivery instead of the machine calculated values. a fault occurs if the loaded database is incorrect.
  • Acción
    Software to be upgraded