Retiro De Equipo (Recall) de GE Magnetic Resonance Imaging System, magnetic rundown unit (MRU)

Según New Zealand Medicines and Medical Devices Safety Authority, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en New Zealand que fue producido por GE Medical Systems LLC.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    18075
  • Fecha de inicio del evento
    2015-02-11
  • País del evento
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: GE Healthcare Limited, 8 Tangihua Street, Auckland 1010
  • Causa
    The mru (magnet rundown unit) is a critical safety component of the mr system that quenches the magnet (i.E., makes it non-magnetic) in emergency situations, such as those that can occur when a ferrous (metal) object is brought into the magnetic field., at certain sites, the mru may not be connected to the magnet. if the mru is disconnected, the mru will not function as expected and will not quench the magnet when the button is pressed. in emergency situations, a disconnected mru could delay removal of a ferrous object from the magnet, potentially resulting in life-threatening injuries. the mru must be connected to the magnet at all times.
  • Acción
    Product to be modified