Retiro De Equipo (Recall) de GE Optima XR220amx & XR220amx with Digital Upgrade X-Ray Systems

Según New Zealand Medicines and Medical Devices Safety Authority, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en New Zealand que fue producido por GE Medical Systems LLC.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    15868
  • Fecha de inicio del evento
    2013-12-13
  • País del evento
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: GE Healthcare Limited, 8 Tangihua Street, Auckland 1010
  • Causa
    Intermittently when a large number of patient exams are retrieved from the his/ris system, the exams when displayed on the mobile x-ray system are not properly displayed. if the user continues to attempt to select one of the exams when this issue has occurred, the system will update the screen and in doing so, potentially select a different exam/patient. if the user does not notice this change and continues to perform the exam, the images will be placed in an incorrect patient file.
  • Acción
    Software to be upgraded

Device

  • Modelo / Serial
    Model: 5555000-5, 5555000-6 and 5421698, Affected: , Software version: All versions