Retiro De Equipo (Recall) de GE Precision and Advantx Legacy/Legacy-D Radiographic and Fluroscopic Systems

Según New Zealand Medicines and Medical Devices Safety Authority, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en New Zealand que fue producido por GE Medical Systems LLC.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    18755
  • Fecha de inicio del evento
    2015-06-29
  • País del evento
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: GE Healthcare Limited, 8 Tangihua Street, Auckland 1010
  • Causa
    The spot film device (sfd) or intelegent digital device (idd) safety mechanism installed on the system may not engage properly at the lower range of sfd/idd travel. this mechanism is installed to hold and prevent the sfd/idd from falling due to a counterweight cable failure. the lower range of travel is 1.1" (28mm) from the lowest point of normal sfd/idd vertical compression. a fall of the sfd/idd could result in an injury to a patient or operator. there have been no reported injuries as a result of this issue.
  • Acción
    Product to be modified