Retiro De Equipo (Recall) de Heater Cooler Unit

Según New Zealand Medicines and Medical Devices Safety Authority, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en New Zealand que fue producido por Maquet Cardiopulmonary AG.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    20582
  • Fecha de inicio del evento
    2016-08-30
  • País del evento
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Maquet Australia Pty Ltd (NZ), 12G Andromeda Cres East Tamaki, Auckland 2013
  • Causa
    Maquet cardiopulmonary have received reports confirming the presence of mycobacterial contamination (and other bacterial contaminants) in the system water of heater and heater/cooler devices, mainly used during cardiothoracic surgeries to warm or cool a patient. although the water in the circuits does not come into direct contact with the patient, there is the potential for contaminated water to enter other parts of the device or transmit bacteria through the air (aerosolize) through the device's exhaust vent into the environment and to the patient.
  • Acción
    Manufacturer to issue advice regarding use