Retiro De Equipo (Recall) de Infusion pumps - Syringe (Swiss)

Según New Zealand Medicines and Medical Devices Safety Authority, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en New Zealand que fue producido por CareFusion Switzerland 317 Sarl.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    22304
  • Fecha de inicio del evento
    2017-11-23
  • País del evento
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: CareFusion New Zealand 313 Limited, Unit 14, 13 Highbrook Drive, East Tamaki, AUCKLAND
  • Causa
    The manufacturer has received a report regarding reoccurring pl3 alarms. a pl3 alarm is an internal malfunction error code, which will result in the pump stopping and alerting the user of an internal failure. if a pump displays a pl3 error message, the pump should be immediately removed from clinical use and sent for servicing by qualified service personnel.
  • Acción
    Manufacturer to issue advice regarding use

Device

  • Modelo / Serial
    Model: , Affected: Product codes with prefix (all variants): 8001, 8002, 8003,8004, and 8005
  • Manufacturer

Manufacturer