Retiro De Equipo (Recall) de Medical Intelligence HexaPOD evo RT System

Según New Zealand Medicines and Medical Devices Safety Authority, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en New Zealand que fue producido por Medical Intelligence Medizintechnik GmbH.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    22989
  • Fecha de inicio del evento
    2018-05-17
  • País del evento
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Elekta Pty Limited - New Zealand Branch, Level 2, Windsor Court , 128-136 Parnell Road, Parnell, AUCKLAND 1052
  • Causa
    It has been noted that upon completion of a 6d workflow, the hexapod evo module may be in a tilted state, if it is moved to the position used for 3d (non-iguide) treatments. in a subsequent 3d workflow, a possible incorrect position of the patient, due to a tilt of the hexapod, may be difficult to detect without verification (e.G. imaging). this description is not relevant for an iguide workflow (6d). the problem occurs only when you switch from a 6d to a 3d workflow without iguide.
  • Acción
    Manufacturer to issue advice regarding use