Retiro De Equipo (Recall) de Philips Allura Systems, Actuator Monitor Ceiling Suspension (MCS)

Según New Zealand Medicines and Medical Devices Safety Authority, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en New Zealand que fue producido por Philips Medical Systems Nederland BV.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    18800
  • Fecha de inicio del evento
    2015-07-06
  • País del evento
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Philips New Zealand Commercial Limited, Level 3, 123 Carlton Gore Road, Newmarket, AUCKLAND 1023
  • Causa
    The manufacturer has received a report of a failing monitor ceiling suspension (mcs) actuator on an allura xper system. the actuator assembly had become detached and the monitor carriage with a flexvision monitor dropped to its lowest position., when the monitor ceiling suspension falls to its lowest position there is a possibility it might collide with a patient or personnel in the room. there is also the posssibility that the monitor ceiling suspension may collide with other equipment in close proximity to the mcs.
  • Acción
    Product to be modified