Retiro De Equipo (Recall) de Philips BrightView XCT

Según New Zealand Medicines and Medical Devices Safety Authority, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en New Zealand que fue producido por Philips Radiation Oncology Systems.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    19646
  • Fecha de inicio del evento
    2016-01-15
  • País del evento
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Philips New Zealand Commercial Limited, Level 3, 123 Carlton Gore Road, Newmarket, AUCKLAND 1023
  • Causa
    Customer complaint indicated that after a software upgrade to version 1.2.3/ 2.5.3 or higher the following issue started to occur. upon completion of a patient study when the operator selected a pre-prorgammed motion (ppm) to remove the patient from the table. the table moved first instead of the detector (as it had done prior to the software upgrade). there is a potential for the patient to contact the collimators if the detctors are placed in close proximity to the patient at the end of an acquisition, followed by acquisition ofanother ppm due to te sequence of motions mentioned above.
  • Acción
    Software to be upgraded

Device

Manufacturer

  • Empresa matriz del fabricante (2017)
  • Source
    NZMMDSA