Retiro De Equipo (Recall) de Philips Heartstart MRx Monitor/Defib

Según New Zealand Medicines and Medical Devices Safety Authority, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en New Zealand que fue producido por Philips Medical Systems (USA).

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    16285
  • Fecha de inicio del evento
    2014-02-24
  • País del evento
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Laerdal (New Zealand) Ltd, Unit O, Cain Commerical Centre, 20 Cain Road, Penrose
  • Causa
    When used in hospital transport and pre-hospital (ems) environments, the connection between the pads/paddles therapy cable (including pads cpr therapy cable, external paddles cable, and internal paddles cable/adapter) and the mrx therapy connection port could experience high levels of stress causing accelerated wear. wear of the connection could prevent the device from sensing that the pads/paddles therapy cable is connected. this wear also could cause the mrx to inappropriately identify the pads therapy cable, external paddles, or internal paddles., pads/paddles therapy cable and therapy connection port wear could pose a risk for patients and/or caregivers.
  • Acción
    Product to be modified

Device

Manufacturer

  • Empresa matriz del fabricante (2017)
  • Source
    NZMMDSA