Retiro De Equipo (Recall) de Philips HeartStart MRx Monitor/Defibrillator

Según New Zealand Medicines and Medical Devices Safety Authority, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en New Zealand que fue producido por Philips Medical Systems (USA).

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    16283
  • Fecha de inicio del evento
    2014-02-24
  • País del evento
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Philips New Zealand Commercial Limited, Level 3, 123 Carlton Gore Road, Newmarket, AUCKLAND 1023
  • Causa
    A component of the mrx processor board may be susceptible to damage from electrostatic discharge (esd). damage to these components can disrupt ecg and spo2 functionality leading to an inability to perform:, " demand mode pacing, " 12 lead ecg acquisition and analysis, " ecg monitoring via leads, " pulse oximetry (spo2), the presence of this condition will be displayed on the ready for use (rfu) indicator as a solid red x with periodic audio chirp., defibrillation, fixed mode pacing, ecg monitoring via pads/paddles, and other monitoring functions are not impacted by this issue.
  • Acción
    Product to be modified